Chapter 5 443
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
Example: MCP:CARR:AUTO 0
MCP:CARR:AUTO?
MCP:CARR -100
MCP:CARR?
5.19.9 Limit Test
Pressing Limit Test turns the testing of the limit line on or off. When Limit Test is set to On,
each offset is compared to its upper and lower offset limit. In those cases where the power
exceeds the limit, a red “F” is placed next to the dBc result to indicate a failure. If there are
any failures, the pass/fail indicator area show “FAIL” in red; if there are none, it shows
“PASS” in green. Any offsets that are in the off state (see
“Offset Freq” on page 438)arenot
measured and their results will not be displayed on screen.
Key Path:
Meas Setup, More
State Saved: Saved in instrument state.
Factory Preset: Off
History: Added with firmware revision A.02.00