Agilent Technologies E4440-90189 Saw User Manual


 
Chapter 5 429
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
5.19 Meas Setup (Multi-Carrier Power—MCP)
If the MCP measurement has been selected in the Measure menu of the Spectrum Analysis
mode, this key displays the MCP measurement setup menu.
The Multi-Carrier Power measurement is a measure of the power in two or more transmit
channels and of the power that leaks into their adjacent transmit channels. The results
reported are similar to the adjacent channel power measurement, but the setup is different
to allow for two or more carriers present.
Key Path: Front-panel key
Dependencies/
Couplings:
Parameter defaults change depending on the Radio Standard selected.
Sweep time is coupled to span, RBW, and VBW if the measurement
method is set to IBW.
History: Added with firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.19.1 Avg Number
Pressing Avg Number to On enables you to specify the number of measurements that will be
averaged when calculating the measurement result. The average will be displayed at the
end of each sweep. Setting Avg Number to Off disables the measurement averaging.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 / Off
Range: 1 through 1000
History: Added with firmware revision A.02.00
Remote Commands:
[:SENSe]:MCPower:AVERage:COUNt <integer>
[:SENSe]:MCPower:AVERage:COUNt?
[:SENSe]:MCPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:AVERage[:STATe]?
Example: MCP:AVER:COUN 10
MCP:AVER:COUN?