Agilent Technologies E4440A Saw User Manual


 
Chapter 3 199
Instrument Functions: M - O
Marker ->
Instrument Functions: M - O
3.3.4 Mkr->Stop
Changes the stop frequency to the frequency of the active marker. The marker stays at this
frequency, so it moves to the right of the display. This function is not available in Zero Span.
Key Path:
Marker ->
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:STOP
Example: CALC:MARK1:STOP sets the stop frequency to the value (or delta value) of
marker 1.
3.3.5 Mkr->Span
Sets the start and stop frequencies to the values of the delta markers. The marker is then
set to normal at the center frequency. Only available in
Delta, Span Pair,andDelta Pair
modes, this function is not available if the marker is off, or in Normal mode or when the
span is Zero Span.
Key Path:
Marker ->
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:SPAN
Remote Command Notes: Select the delta marker mode with
:CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example: CALC:MARK2:DELT:SPAN sets the start and stop frequencies to the values of
the delta markers of marker 2.
3.3.6 Mkr->CF
Sets the delta marker to the center frequency. Only available in Delta, Span Pair,andDelta
Pair
modes, this function is not available if the marker is off, or in Normal mode or when the
span is
Zero Span.
Key Path:
Marker ->
History: Added with firmware revision A.02.00
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:CENTer
Remote Command Notes: Select the delta marker mode with
:CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example: CALC:MARK2:DELT:CENT sets the center frequency to the value of the delta
marker center frequency of marker 2.