Agilent Technologies E4440A Saw User Manual


 
Chapter 5 441
One-Button Measurement Functions
Meas Setup (Multi-Carrier Power—MCP)
One-Button Measurement Functions
5.19.5 Carrier Result
Press Carrier Result to select the result you wish to display on the last line of the carrier
power results list except when:
the carrier result number 4 (the first 4 carrier power results are displayed)
the carrier result number 9 (the last 4 carrier power results are displayed)
Key Path:
Meas Setup
Key Notes: This key is only available when Meas Control, Measure (Single) is selected.
State Saved: Not saved.
Factory Preset: 1
Step Key
Increment: 1
Range: 1 to number of carriers.
History Added with firmware revision A.03.00
Remote Command:
There is no equivalent remote command.
5.19.6 Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, which is its
freedom from errors due to input compression.
NOTE There will always be a minimum of 2 dB of attenuation set to reduce the
potential errors due to VSWR.
Key Path:
Meas Setup
State Saved: Not saved.
History: Added with firmware revision A.02.00
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example: POW:RANG:AUTO ONCE
5.19.7 Method
Enables you to set the measurement method to either the integration bandwidth method
(IBW) or the resolution bandwidth method (RBW). The resolution bandwidth method is