460 Chapter5
One-Button Measurement Functions
Meas Setup (Spectrum Emissions Mask—SEM)
One-Button Measurement Functions
Remote Command:
[:SENSe]:SEMask:SWEeptime <number>
[:SENSe]:SEMask:SWEeptime?
[:SENSe]:SEMask:SWEeptime:AUTO OFF|ON|0|1
[:SENSe]:SEMask:SWEeptime:AUTO?
Example: SEM:SWE 4 s
SEM:SWE?
SEM:SWE:AUTO 1
SEM:SWE:AUTO?
5.22.3.4 Res BW
Specifies the resolution bandwidth used in measuring and the power in the main channel.
Key Path:
Meas Setup, Ref Channel
Factory Preset: 100 kHz/On
Range: Full Range provided by Base Instrument Hardware.
History: Remote command updated with firmware revision A.03.00
Remote Command:
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]?
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:AUTO?
Remote Command Notes: If an unavailable bandwidth is entered, the closest available
bandwidth is used. Refer to
“Res BW” on page 83.
Example: SEM:BAND 4 MHz
SEM:BWID 4 MHz
SEM:BAND?
SEM:BWID?
SEM:BAND:AUTO 1
SEM:BWID:AUTO 1
SEM:BAND:AUTO?
SEM:BWID:AUTO?