Agilent Technologies E4440A Saw User Manual


 
Chapter 5 485
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
5.27 Meas Setup (Spurious Emissions)
When the spurious emissions measurement has been selected in the Measure menu of the
Spectrum Analysis mode, this key displays the appropriate measurement setup menu.
The spurious emissions measurement identifies and determines the power level of
spurious emissions in certain frequency bands.
Key Path: Front-panel key
History: Added with firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.27.1 Avg Number
Press Avg Number to On to specify the number of measurements that will be averaged when
calculating the measurement result. The average will be displayed at the end of each
sweep.
Off disables the measurement averaging.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 averages / Off
Range: 1 to 1000
Remote Command:
[:SENSe]:SPURious:AVERage:COUNt <integer>
[:SENSe]:SPURious:AVERage:COUNt?
[:SENSe]:SPURious:AVERage[:STATe] OFF|ON|0|1 turns the averaging on or off.
[:SENSe]:SPURious:AVERage[:STATe]?
Example: SPUR:AVER:COUN 10
SPUR:AVER OFF