Agilent Technologies E4440A Saw User Manual


 
Chapter 8 579
Using the STATus System
STATus Subsystem
Using the STATus System
Condition Register will set bits in the Questionable Integrity Event
register, which also sets the Integrity Summary bit (bit 9) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Integrity Event Query
:STATus:QUEStionable:INTegrity[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Integrity Event register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Integrity Negative Transition
:STATus:QUEStionable:INTegrity:NTRansition <number>
:STATus:QUEStionable:INTegrity:NTRansition?
This command determines which bits in the Questionable Integrity
Condition register will set the corresponding bit in the Questionable
Integrity Event register when the condition register bit has a negative
transition (1 to 0)
Thevariable<number>isthesumofthedecimalvaluesofthebitsthat
you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 0
Range: 0 to 32767
Questionable Integrity Positive Transition
:STATus:QUEStionable:INTegrity:PTRansition <number>
:STATus:QUEStionable:INTegrity:PTRansition?