356 Chapter5
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
Default Terminator: Hz
Knob Increment: Span/50 if non-zero span.
RBW/100 if zero span.
Step Key
Increment: RBW if
CF Step is set to Auto with zero span.
Span/10 if
CF Step is set to Auto with non-zero span.
CF Step if
CF Step is set to Manual.
Range: > 0 Hz to 45 MHz
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:OFFSet:LIST[:FREQuency]
<f_offset>,<f_offset>,<f_offset>,<f_offset>,<f_offset>,<f_offset>
[:SENSe]:ACPower:OFFSet:LIST[:FREQuency]?
[:SENSe]:ACPower:OFFSet:LIST:STATE
OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:ACPower:OFFSet:LIST:STATE?
Remote Command Notes: This command, along with commands
[:SENSe]:ACPower:OFFSet:LIST:BANDwidth|BWIDth[:INTegration]
and [:SENSe]:ACPower:OFFSet:LIST:STATe, are used to set the
parameters for all of the offsets, up to six. The following table shows the
values of these parameters if no radio standard (
None) is selected.
Offset State Frequency Integ. BW
A ON 3 MHz 2 MHz
BOFF0 2MHz
COFF0 2MHz
DOFF0 2MHz
EOFF0 2MHz
FOFF0 2MHz