Chapter 5 399
One-Button Measurement Functions
Meas Setup (Channel Power—CHP)
One-Button Measurement Functions
5.14 Meas Setup (Channel Power—CHP)
When the channel power measurement has been selected in the Measure key menu of the
Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
The Channel Power measurement measures the power and power spectral density in the
channel bandwidth that you specify. One marker pair on the display indicates the edges of
the channel bandwidth. The center frequency, reference level, and channel bandwidth
must be set by the user.
Key Path: Front-panel key
Dependencies/
Couplings: Menu changes depending on the Mode and Measurement selected.
State Saved: No save
History: Added with firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.14.1 Avg Number
To specify the number of measurement averages used when calculating the measurement
result set
Avg Number to On. The average will be displayed at the end of each sweep. Setting
Avg Number to Off disables measurement averaging.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 averages/ Off
Knob Increment: 1
Step Key
Increment: 1
Range: 1 to 1000
Remote Command:
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?