Agilent Technologies E4440A Saw User Manual


 
496 Chapter5
One-Button Measurement Functions
Meas Setup (Spurious Emissions)
One-Button Measurement Functions
5.27.3.9 Peak Excursn
Peak Excursn sets the minimum amplitude variation of signals that can be identified as
peaks. For example, if a value of 6 dB is selected, peaks that rise and fall more than the 6
dB above the peak threshold value are identified.
Key Path:
Meas Setup, Range Table
Factory Preset: See “Range Table” on page 486.
Knob Increment: 1 dB
Step Key
Increment: 1 dB
Range: 0.0 dB to 100.0 dB
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion <integer>
The query for this parameter will always return 20 values.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:EXCursion?
5.27.3.10 Pk Threshold
Pk Threshold sets the minimum amplitude of signals that can be identified as peaks. For
example, if a value of -90 dBm is selected, only peaks that rise and fall more than the peak
excursion value which is above -90 dBm are identified.
Key Path:
Meas Setup, Range Table
Factory Preset: See “Range Table” on page 486.
Knob Increment: 1 dB
Step Key
Increment: 1 dB
Range: –100 dB to 0.0 dB
Remote Command:
This parameter can receive up to 20 values. The location in the list sent corresponds to the
range the value is associated with. Missing values are not permitted, for example, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values will
remain as they were.
[:SENSe]:SPURious[:RANGe][:LIST]:PEAK:THReshold <integer>