NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
register is cleared.
Key Type: There is no equivalent front panel key.
Questionable Calibration Negative Transition
:STATus:QUEStionable:CALibration:NTRansition <number>
:STATus:QUEStionable:CALibration:NTRansition?
This command determines which bits in the Questionable Calibration
Condition register will set the corresponding bit in the Questionable
Calibration Event register when the condition register bit has a
negative transition (1 to 0). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 0
Range: 0 to 32767
Questionable Calibration Positive Transition
:STATus:QUEStionable:CALibration:PTRansition <number>
:STATus:QUEStionable:CALibration:PTRansition?
This command determines which bits in the Questionable Calibration
Condition register will set the corresponding bit in the Questionable
Calibration Event register when the condition register bit has a positive
transition (0 to 1). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Frequency Register
Questionable Frequency Condition
:STATus:QUEStionable:FREQuency:CONDition?