414 Chapter5
One-Button Measurement Functions
Meas Setup (Harmonic Distortion)
One-Button Measurement Functions
Factory Preset: For each range, the following settings are selected: Range State(On), Zero
Span
, Res BW Mode(Man), Video BW(Man) is set to 2 kHz, and Sweep Time
Mode
(Auto) is set to 10 ms. The table below defines the other default
parameter settings for the measurement.
History: Added with firmware revision A.03.00
Remote Command:
There is no equivalent remote command.
5.16.6.1 Range
Selects a range and updates the values on the other
Range Table keys so that they reflect the
settings for the selected range. If
Range is set to On it is used as part of the measurement;
when set to
Off it is excluded. A range is made up of the next parameters.
Key Path:
Meas Setup, Range Table
Factory Preset: See Table 5-1 on page 414.
Range: Off|On
Table 5-1 Range Table Default Settings
Range Center
Freq (GHz)
Res BW
(kHz)
1 1.000 100.0
2 2.000 200.0
3 3.000 300.0
4 4.000 430.0
5 5.000 510.0
6 6.000 620.0
7 7.000 750.0
8 8.000 820.0
9 9.000 910.0
10 10.00 1000.0