Chapter 5 359
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions
Remote Command:
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA
<rel_power>,<rel_power>,<rel_power>,<rel_power>,<rel_power>,<rel_power>
:CALCulate:ACPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA?
Example: CALC:ACP:OFFS:LIST:LIM:POS:DATA 10,10,10,10,10,10
CALC:ACP:OFFS:LIST:LIM:POS:DATA?
5.6.4.5 Neg Offset Limit
Enables you to set the upper limit for the lower segment of the specified offset pair.
Key Path:
Meas Setup, Offset Setup
State Saved: Saved in instrument state.
Factory Preset: 0.00 dB
Default Terminator: dB
Range: –200 dB to 200 dB
History: Added with firmware revision A.02.00.
Remote Command:
:CALCulate:ACPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA
<rel_power>,<rel_power>,<rel_power>,<rel_power>,<rel_power>,<rel_power>
:CALCulate:ACPower:OFFSet:LIST:LIMit:NEGative[:UPPer]:DATA?
Example: CALC:ACP:OFFS:LIST:LIM:NEG:DATA 50,50,50,50,50,50
CALC:ACP:OFFS:LIST:LIM:NEG:DATA?
5.6.5 Meas Type
Press Meas Type to specify the reference for the measurement, either Total Pwr Ref or PSD Ref.
Relative values can be displayed referenced to either the total power (Total Pwr Ref)orthe
power spectral density (PSD Ref) measured in the main channel.
Key Path:
Meas Setup
State Saved: Saved in instrument state.
Factory Preset: Total Pwr Ref
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:TYPE PSDRef|TPRef