518 Chapter6
Programming Fundamentals
Improving Measurement Speed
Programming Fundamentals
Table 6-1 GSM Parameters for 1 Slot/Frame Measurement Requirements
Resolution
Bandwidth
Filter
Type
Decimation Aperture Start Length Repeat
500 or 300
kHz
Flat or
Gaussian
4 or 1 dependent
on settings
24 µsec
a
526 µsec
a
576.9 µsec
a
500 kHz Gaussian 1 0.2 µsec 124 2630 2884.6
500 kHz Gaussian 4 0.8 µsec 31 657 721.15
500 kHz Flat 1 0.4 µsec 61 1315 1442.3
500 kHz Flat 4 1.6 µsec 15 329 360.575
300 kHz Gaussian 1 0.2667 µsec 90 1972 2163.1
300 kHz Gaussian 4 1.07 µsec 22 492 539.16
300 kHz Flat 1 0.6667 µsec 36 789 865.31
300 kHz Flat 4 2.667 µsec 9 197 216.33
a. The use of time values is only allowed in firmware versions of A.02.00 and later.