Agilent Technologies E4440A Saw User Manual


 
Chapter 2 67
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.0.10.3 Freq Interp
Allows you to determine how trace values are computed between points in a correction
table. If the linear mode is selected, a straight line is used between points in the correction
table (for a linear frequency scale and for a decibel amplitude scale). If the logarithmic
mode is selected, frequency values between points are computed by first taking the
logarithm of both table values and the intermediate value, while using decibel amplitude
values.
Key Path:
AMPLITUDE / Y Scale, More, Corrections
State Saved: Not saved in instrument state
Factory Preset: Not affected by Factory Preset. Will be set to linear by Restore Factory
Defaults.
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4:X:SPACing LINear|LOGarithmic
Example: CORR:CSET4:X:SPAC LOG
2.0.10.4 Delete All Corrections
Allows you to delete all amplitude-correction sets.
Key Path:
AMPLITUDE / Y Scale, More, Corrections
Dependencies/
Couplings: If corrections are on, corrections are turned off.
Remote Command:
[:SENSe]:CORRection:CSET:ALL:DELete
Example: CORR:CSET:ALL:DEL