Agilent Technologies E4440A Saw User Manual


 
582 Chapter8
Using the STATus System
STATus Subsystem
Using the STATus System
Questionable Integrity Uncalibrated Register
Questionable Integrity Uncalibrated Condition
:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Integrity Uncalibrated Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Key Type: There is no equivalent front panel key.
Questionable Integrity Uncalibrated Enable
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle?
This command determines which bits in the Questionable Integrity
Uncalibrated Condition Register will set bits in the Questionable
Integrity Uncalibrated Event register, which also sets the Data
Uncalibrated Summary bit (bit 3) in the Questionable Integrity
Register. The variable <number> is the sum of the decimal values of the
bits you want to enable.
Key Type: There is no equivalent front panel key.
Factory Preset: 32767 (all 1’s)
Range: 0 to 32767
Questionable Integrity Uncalibrated Event Query
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Integrity Uncalibrated Event register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
register is cleared.
Key Type: There is no equivalent front panel key.