Philips 200 Work Light User Manual


 
Trilogy200 clinical manual
112
Alarm Priority Audible Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
High
Respiratory
Rate
High
• • • • •
Red ashing button;
“High Respiratory
Rate” message
Operates User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
Low Respiratory
Rate
High
• • • • •
Red ashing button;
“Low Respiratory
Rate” message
Operates User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
High
Inspiratory
Pressure
Escalates
from
Audible
indicator
to Medium
priority and
then High
priority
(for rst two
consecutive
occurrences)
• • • (for 3rd
consecutive
occurrence)
• • • • •
(for 10th
consecutive
occurrence)
Peak Pressure symbol
turns red. When
condition rst occurs,
a beep will sound.
When condition
occurs for third time,
button ashes yellow
and yellow “High
Inspiratory Pressure”
message appears.
When condition
occurs for 10th
time, button ashes
red and red “High
Inspiratory Pressure”
message appears.
Operates User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. Make
sure tubing is not pinched or
blocked. If alarm continues,
contact your health care
professional.
Clinician: Verify patient
status. If problem continues,
have device serviced.