Trilogy200 clinical manual
116
Alarm Priority Audible Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Keypad Stuck Low
• •
Solid yellow button;
“Keypad Stuck”
message.
Operates User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Check the keys to determine
if they are lodged in the case.
If alarm continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: Check keys
to determine if they are
lodged in the case. If alarm
continues, place patient
on alternate source of
ventilation and have device
serviced.
Battery
Discharging
Stopped Due to
Temperature
Info
•
“Batt Discharge
Stopped – Temp.”
message
Operates User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
Left (Reset) button to reset
alarm. Move device to a
cooler location. Make sure
device is not close to a heat
source. Make sure cooling
vents are not blocked. If
alarm is detected for internal
battery and continues, place
patient on alternate source of
ventilation and contact your
health care professional.
Clinician: If alarm is detected
on the internal battery and
continues after the measures
above have been taken, place
patient on alternate source of
ventilation and have device
serviced.