Chapter 11 Technical Specications
151
Environmental
Operating Storage
Temperature 41° F to 104° F (5° C to 40° C) -4° F to 140° F (-20° C to 60° C)
Relative Humidity 15 to 95% (non-condensing) 15 to 95% (non-condensing)
Atmospheric Pressure 110 kPa to 60 kPa N/A
The operating range for the charging of the internal and detachable batteries is 10° C to 30° C. The internal and detachable
batteries will power the ventilator for the full operating range of 5° C to 40° C.
Accuracies stated in this manual are based on specic environmental conditions. For stated accuracy, the environmental
conditions are: Temperature: 20-30° C; Humidity: 50% relative; Altitude: nominally 380 meters.
Physical
Dimensions: 6.6” L x 11.2” W x 9.3” H (16.68 cm L x 28.45 cm W x 23.52 cm H)
Weight: Approximately 11 lbs (5 kg) with the detachable battery installed
Standards Compliance
This device is designed to conform to the following standards:
IEC 60601-1: Medical electrical equipment - Part 1: General requirements for safety
IEC 60601-1-2: General requirements for safety - Collateral standard: Electromagnetic
compatibility - Requirements and tests
ISO 10651-2-Lung ventilators for medical use -- Particular requirements for basic safety and
essential performance -- Part 2: Home care ventilators for ventilator-dependent patients
RTCA-D0160F section 21, category M; Emission of Radio Frequency Energy
Trilogy200
clinical manual
11. Technical Specications