Chapter 4 Ventilator Setup
61
Connect Supplemental Oxygen (Optional)
To attach supplemental oxygen to the device:
1. Connect the oxygen tubing to the Oxygen Inlet Quick Connect.
Then, connect the Quick Connect to the Oxygen Inlet on the
back of the ventilator.
Oxygen Inlet
Oxygen Source
Tubing
Oxygen Inlet
Quick Connec
t
2. Make sure you press the Oxygen Inlet Quick Connect rmly into
the inlet. You should hear a click, and the latch at the top of the
inlet will pop up.
This device is only compatible with a low ow oxygen source
providing up to 15 l/min.
Attaching the O
2
Tubing
Press Tubing in and Latch
Pops Up
CAUTION
Do not connect an unregulated
or high pressure oxygen source