Trilogy200 clinical manual
80
Note: The Ramp Start Pressure
setting will not display if the
Ramp Length is set to O or if
the CPAP pressure is set to
4 cm H
2
O.
Note: Ramp Start Pressure is
less than or equal to CPAP - 1 cm
H
2
O in CPAP mode.
Note: In CPAP mode, Flex is only
available when CPAP is greater
than 4 cm H
2
O.
Note: In S mode, Flex is only
available when EPAP is greater
than or equal to 4 cm H
2
O and
IPAP is less than or equal to
25 cm H
2
O.
Note: Flex is only available
when Auto-Trak is enabled.
Note: Ramp is not available in
the Passive Circuit when IPAP =
EPAP = 4 cm H
2
O or when IPAP
Min = EPAP = 4 cm H
2
O.
Note: Ramp Start Pressure is
less than or equal to EPAP - 1 cm
H
2
O in S, S/T, T, and PC modes.
Note: Ramp Start Pressure is
greater than or equal to 0 cm
H
2
O when the circuit type is
Active PAP or Active Flow in S,
S/T, T, and PC modes.
Note: Ramp Start Pressure
is greater than or equal to 4
cm H
2
O when the circuit type
is Passive in S, S/T, T, and PC
modes.
Note: AVAPS is only available if
the Passive Circuit is selected.
7. Ramp Start Pressure
You can increase or decrease the ramp start pressure in
increments of 1 from 4 cm H
2
O to the CPAP pressure setting. The
patient also has access to this setting, unless the ramp length is
set to O.
8. Flex
You can enable or disable the Flex setting. O disables the
setting and prevents the patient from using Flex. To enable the
setting, set Flex to 1, 2, or 3. The patient also has access to this
setting, if Flex is enabled. The Flex feature is not available when
using an active circuit type.
Spontaneous (S) Mode
The following settings, described in the CPAP mode section of this
chapter, also are available in S mode:
Flex
Trigger Type
Flow Trigger Sensitivity
Flow Cycle Sensitivity
Ramp Length
Ramp Start Pressure
In addition to those settings, the settings below are also available in
S mode.
1. AVAPS
AVAPS is only available if Flex is not enabled.
You can disable AVAPS by selecting O, or you can enable AVAPS
by selecting On. If you select O, the IPAP setting displays. If you
select On, the IPAP Max Pressure and IPAP Min Pressure display.