Low Minute Ventilation
This setting enables or disables the Low Minute Ventilation alarm. The
alarm activates when the calculated minute ventilation is less than or
equal to this setting. You can choose O to disable this alarm, or you
can increase or decrease the setting from 0.1 l/min to 99 l/min.
High Minute Ventilation
This setting enables or disables the High Minute Ventilation alarm.
The alarm activates when the calculated minute ventilation reaches
or exceeds this setting. You can choose O to disable this alarm, or
you can increase or decrease the setting from 1 l/min to 99 l/min
in 1 l/min increments. It cannot be set lower than the Low Minute
Ventilation setting except to be turned o.
Low Respiratory Rate
This setting enables or disables the Low Respiratory Rate alarm. The
alarm activates when the measured respiratory rate is less than or
equal to this setting. You can choose O to disable this alarm, or you
can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM
increments. It cannot be set higher than the High Respiratory Rate.
High Respiratory Rate
This setting enables or disables the High Respiratory Rate alarm.
The alarm activates when the measured respiratory rate reaches or
exceeds this setting. You can choose O to disable this alarm, or you
can increase or decrease the setting from 4 BPM to 80 BPM in 1 BPM
increments. It cannot be set lower than the Low Respiratory Rate
except to be turned o.
Note: You can increase or
decrease the Low Minute
Ventilation setting in
0.1 l/min increments from
0.1 to 0.99 l/min and 1 l/min
increments from 1 to 99 l/min.