Chapter 3 Modes, Features, and Alarms
25
Pressure Control (PC) Mode
In Pressure Control (PC) mode, the device delivers bi-level pressure
support. This mode delivers Assist and Mandatory breaths. This
mode is identical to S/T mode, except that all breaths have a xed
inspiratory time.
= Mandatory Breath
= Assist Breath
IPAP = 26 cm H
2
O
EPAP = 6 cm H
2
O
Rate = 10 BPM
PS = IPAP-EPAP = 20 cm H
2
O
3 3
3
3
2
2
6
10
20
26
Enter PC mode
5 10 15 20
PRESSURE
TIME
6 seconds
6 seconds 6 seconds
Pressure Control - Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) Mode
The Pressure Control – Synchronized Intermittent Mandatory
Ventilation (PC-SIMV) mode provides Spontaneous, Assist, and
Mandatory breaths. This mode uses a time window to decide
what type of breaths should be delivered. This time window is the
maximum time between breaths according to the set Breath Rate.
When you enter PC-SIMV mode, the time window is started. If
the patient does not trigger a breath, the ventilator will provide
a Mandatory breath when the time expires and then begin the
process again. Mandatory breaths deliver the Pressure setting during
inhalation and a lower Peak End Expiratory Pressure (PEEP)
during exhalation.
PC Mode