Trilogy200 clinical manual
48
Check External Battery Info Message
This info message occurs when a bad connection exists to the
external battery or the external battery failed. The device continues
to operate using power from the detachable battery, if available, or
the internal battery.
Battery Depleted Info Message
This info message occurs when the aected battery is fully depleted.
The device continues to operate using the next available power source.
External Battery Disconnected Info Message
This info message occurs when the external battery power source is
lost and the device has switched to an alternate power source (either
a detachable battery, if connected, or the internal battery if no other
source is available). If external battery power returns, the ventilator
will beep, but no message will appear on the display.
Detachable Battery Disconnected Info Message
This info message occurs when the detachable battery power source
is lost and the device has switched to an alternate power source (the
internal battery if no other source is available). If detachable battery
power returns, the ventilator will beep, but no message will appear
on the display.
Start On Battery Info Message
This info message indicates that the ventilator has started on battery
power and no AC power is available. The device operator should
verify that this is what is wanted.
Card Error Info Message
This info message occurs when an unusable SD card is inserted into
the ventilator. The device continues to operate but data cannot be
logged onto the SD card.