Philips 200 Work Light User Manual


 
Trilogy200 clinical manual
26
During the time window, if patient eort is detected, either a
Spontaneous or an Assist breath is delivered for the rst eort. If the
last breath delivered was Mandatory, then a Spontaneous breath is
delivered for the rst patient eort of the window. If the last breath
delivered was an Assist or Spontaneous breath, then an Assist breath
is delivered for the rst patient eort of the window. Assist breaths
deliver the Pressure setting during inhalation and the PEEP setting
during exhalation.
During the time window, if patient eort continues to be detected
(after the rst patient triggered breath), Spontaneous breaths are
delivered for the remainder of the window. The gure below provides
example breath patterns in PC-SIMV mode.
12 seconds
2
3
3
111
Pressure = 26 cm H
2
O
Pressure Support = 10 cm H
2
O
PEEP = 6 cm H
2
O
Rate = 5 BPM
= Spontaneous
Breath
= Mandatory Breath
= Assist Breath
= Mandatory Window
= Spontaneous Window
1
2
5 10 15 20 25 30 35 40 45
6
10
20
26
Enter PC-SIMV mode
M
M MM M
S
S S
PRESSURE
TIME
12 seconds 12 seconds 12 seconds
2
Sample Breath Patterns in
PC-SIMV Mode