34 Series N6700 User’s Guide
Effect on Specifications
Specifications for outputs operating in series can be obtained from
the specifications for single outputs. Most specifications are
expressed as a constant or a percentage (or ppm) plus a constant.
For series operation, the percentage portion remains unchanged
while constant portions or any constants are changed as indicated.
Voltage All series specifications referring to voltage are twice the single output
specification except for programming resolution, which is the same as for a
single output.
Current All series specifications referring to current are the same as for a single
output except for CC load effect, CC load cross regulation, CC source effect,
and CC short term drift. These are twice the current programming accuracy
(including the percentage portion) at all operating points.
Load Transient
Recovery Time
Load transient specifications are typically twice the single output.