Agilent Technologies N6700 Welding System User Manual


 
Appendix A Specifications
76 Series N6700 User’s Guide
Agilent Models N6751A/N6752A, N6753A/N6754A, N6761A/N6762A
Performance Specifications
N6751A / N6752A N6753A N6754A N6761A / N6762A
DC Output Ratings:
Voltage 50 V 20 V 60 V 50 V
Current (derated 1% per °C above 40°C) 5 A / 10A 50 A 20 A 1.5 A / 3 A
Power 50 W / 100 W 300 W 300 W 50 W / 100 W
Output Ripple and Noise (PARD):
(from 20 Hz – 20 MHz)
CV peak-to-peak 4.5 mV 5 mV 6 mV 4.5 mV
CV rms 0.35 mV 1 mV 1 mV 0.35 mV
Load Effect (Regulation):
(for any output load change, with a maximum load-lead drop of 1V/lead)
Voltage 2 mV 2 mV 2 mV 0.5 mV
Current 2 mA 12 mA 5 mA
30 μA (@ 0 – 7 V)
65 μA (@ 7 – 50 V)
Source Effect (Regulation):
Voltage 1 mV 0.5 mV 1.2 mV 0.5 mV
Current 1 mA 5 mA 2 mA
30 μA
Programming Accuracy:
(@ 23 °C ±5 °C after 30 min. warm-up. Applies from min. to max. programming range)
Voltage high range 0.06% + 19 mV
0.06% + 10 mV 0.06% + 25 mV
0.016% + 6 mV
Voltage low range ( 5.5 V)
N/A N/A N/A 0.016% + 1.5 mV
Current high range 0.1% + 20 mA 0.1% + 30 mA 0.1% + 12 mA
0.04% + 200 μA
Current low range ( 100mA, @ 0 - 7 V)
( 100mA, @ 7 - 50 V)
N/A
N/A
N/A
N/A
N/A
N/A
0.04% + 15 μA
0.04% + 55 μA
Measurement Accuracy:
(at 23 °C ±5 °C)
Voltage high range 0.05% + 20 mV
0.05% + 10 mV 0.05% + 25 mV
0.016% + 6 mV
Voltage low range ( 5.5 V)
N/A N/A N/A 0.016% + 1.5 mV
Current high range 0.1% + 4 mA 0.1% + 30 mA 0.1% + 8 mA
0.04% + 160 μA
Current low range ( 100mA, @ 0 - 7 V)
( 100mA, @ 7 - 50 V)
N/A
N/A
N/A
N/A
N/A
N/A
0.03% + 15 μA
NOTE 1
0.03% + 55 μA
200 μA current range (Option 2UA)
N/A N/A N/A 0.5% + 100 nA
Load Transient Recovery Time: (time to recover to within the settling band following a load change
- from 60% to 100% and from 100% to 60% of full load for models N6751A & N6761A
- from 50% to 100% and from 100% to 50% of full load for models N6752A, N6762A, N6753A, & N6754A.)
Voltage settling band ± 75 mV
NOTE 2
± 30 mV
NOTE 3
± 90 mV
NOTE 4
± 75 mV
Time
< 100 μs < 100 μs < 100 μs < 100 μs
1
Applies when measuring 4096 data points (SENSe:SWEep:POINts = 4096).
2
When relay option 761 is installed, the settling band is ±125 mV for Model N6752A.
2
When relay option 760 or 761 is installed, the settling band is ±200 mV for Model N6753A.
4
When relay option 760 or 761 is installed, the settling band is ±350 mV for Model N6754A.