Appendix A Specifications
84 Series N6700 User’s Guide
Agilent Models N6773A - N6776A
Performance Specifications
N6773A N6774A N6775A N6776A
DC Output Ratings:
Voltage 20 V 35 V 60 V 100 V
Current
NOTE 1
15 A
NOTE 2
8.5 A 5 A 3 A
Power 300 W 300W 300 W 300 W
Output Ripple and Noise (PARD):
(from 20 Hz – 20 MHz)
CV peak-to- peak 20 mV 22 mV 35 mV 45 mV
CV rms 3 mV 5 mV 9 mV 18 mV
Load Effect (Regulation):
(with output change from no load to full load, up to a maximum load-lead drop of 1V/lead)
Voltage 13 mV 16 mV 24 mV 45 mV
Current 6 mA 6 mA 6 mA 6 mA
Source Effect (Regulation):
Voltage 2 mV 4 mV 6 mV 10 mV
Current
1 mA 1 mA 1 mA 1 mA
Programming Accuracy:
(@ 23 °C ±5 °C after 30 minute warm-up. Applies from minimum to maximum programming range)
Voltage 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% +100 mV
Current
0.15% + 60 mA 0.15% + 60 mA 0.15% + 60 mA 0.15% + 30 mA
Measurement Accuracy:
(at 23 °C ±5 °C)
Voltage 0.1% + 20 mV 0.1% + 35 mV 0.1% + 60 mV 0.1% +100 mV
Current
0.15% + 15 mA 0.15% + 12 mA 0.15% + 12 mA 0.15% + 6 mA
Load Transient Recovery Time:
(time to recover to within the settling band following a load change
from 50% to 100% and from 100% to 50% of full load.)
Voltage settling band ± 0.3 V
NOTE 3
± 0.3 V
NOTE 3
± 0.5 V ± 1.0 V
Time
< 250 µs < 250 µs < 250 µs < 250 µs
1
Output current is derated 1% per °C above 40°C.
2
When relay Option 760 is installed, the maximum output current will be limited to 10 A.
3
When relay Option 760 or 761 is installed, the settling band is ±0.35 V.