Tektronix TDS5000B Series Sander User Manual


 
Brief Procedures
TDS5000B Series Specifications and Performance Verification
2-7
STOP. In the following procedures, do not make changes to front-panel settings
that are not specifically called out. Each verification procedure requires that you
to set the oscilloscope to certain default settings before verifying functions. If
you make changes to settings other than those called out in a procedure, you
may obtain invalid results. In such cases, repeat the procedure starting over from
step 1.
If you are instructed to press a front-panel or screen button, the button may
already be selected (its label will be highlighted). If this is the case, it is not
necessary to press the button.
Equipment
required
One 10X oscilloscope probe, such as Tektronix P5050
Prerequisites None
1. Initialize the oscilloscope: Push the front-panel DEFAULT SETUP button.
2. Hook up the signal source: Connect the probe to the probe compensation
connector and the channel input that you want to test (beginning with CH 1)
as shown in Figure 2--2.
TDS5000B Series oscilloscope
Probe from PROBE
COMPENSATION
outputto CH 1input
Figure 2- 2: Universal test hookup for functional tests - CH 1 shown
3. Turn off all channels:If any of the front-panel channel buttons are lighted,
push those buttons to turn off the displayed channels. See Figure 2--3.
Verify A ll Input Channels