Performance Tests
2-48
TDS5000B Series Specifications and Performance Verification
H Click Upper Limit; then use the multipurpose knob to slowly
decrease the Upper Limit readout until triggering stops.
H CHECK that the Upper Limit readout, after the oscilloscope loses
triggering, is within 1.9 sto2.1s, inclusive.
H Enter the time in the test record.
4. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.
Equipment
required
One sine wave generator (Item 13)
One precision 50 coaxial cable (Item 4)
One 50 termination (Item 4)
One 10X attenuator (Item 1)
One 5X attenuator (Item 2)
Prerequisites Read Prerequisit es on page 2--17 and foot note warnings on page 2--19.
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope: Press the DEFAULT SETUP button.
b. Modify the initialized front-panel control settings:
H Set the Horizontal SCALE to 20 ns (25 ns for TDS5054B E).
H Press the Trigger MODE button to toggle it to Normal.
H From the toolbar, click Horiz, and select the Acquisition tab.
H Click Average, and set the number of averages to 16.
c. Hook up the test-signal source:
H Connect one end of a precision 50 coaxial cable to the signal
output of the generator. Connect the other end of the coaxial cable to
CH 1 through a 50 termination. See Figure 2--20.
Check Sensitivity, Edge
Trigger, DC Coupled