Tektronix TDS5000B Series Sander User Manual


 
Performance Tests
TDS5000B Series Specifications and Performance Verification
2-33
i. Check remaining vertical scale settings against limits (optional):
H If desired, finish checking the remaining vertical scale settings for
the channel under test by repeating substeps d through h for each of
the remaining scale settings listed in Table 2--3 for the channel under
test.
H Before doing substep f, click the Clear button to remove the
previous channel measurements.
H When doing substep f, skip the subparts that turn on the CH<x>
Pk-Pk measurement until you check a new channel.
H Install/remove attenuators between the generator leveling head and
the channel input as needed to obtain the six division reference
signals listed in the table.
j. Test all channels: Repeat substeps a through h for all channels.
3. Disconnect the hookup: Disconnect the test hook up from the input
connector of the channel last tested.
Equipment
required
One sine wave generator (Item 13)
Three precisi on 50 coaxial cables (Item 4)
One power divider (Item 15) or dual input coupler (item 7)
3 SMA female to female adapter connector (Item 16)
3 SMA male-to-female BNC adapter connector (Item 17)
Prerequisites Read Prerequisit es on page 2--17 and foot note warnings on page 2--19.
STOP. Do not use the vertical position knob to reposition any channel while
doing this check. To do so invalidates the test.
1. Install the test hookup and preset the instrument controls:
a. Initialize the front panel: Press the DEFAULT SETUP button.
b. Modify the initialized front-panel control settings:
H Do not adjust the vertical position of any channel during this
procedure.
H From the toolbar, click the Vert button.
H Set the termination of each channel to 50 by selecting each
channel tab and clicking the Termination 50 button.
Check Delay Bet ween
Channels