Tektronix TDS5000B Series Sander User Manual


 
Performance Tests
TDS5000B Series Specifications and Performance Verification
2-19
Table 2-1: Test equipment (Cont.)
Item number and
description
PurposeExampleMinimum requirements
16. Adapter (three required) SMA female-to-female Tektronix part number
015-1012-00
Checking the delay between
channels
17. Adapter (three required) SMA male-to-female BNC Tektronix part number
015-1018-00
Checking the delay between
channels
18. Pulse Generator 250 MHz, 1 ns rise time, 5 V out Wavetek 9500
1,2
UsedtoTestDeltaTime
Measurement Accuracy
19. Cable, Coaxi al
(two required)
50 , 20 in, male-to-male SMA
connectors
Tektronix part number
174-1427-00
UsedtoTestDeltaTime
Measurement Accuracy
20. Adapter (four required) Male N-to-female BNC Tektronix part number
103-0045-00
Checking Analog Bandwidth
21. Adapter Female N-to-male BNC Tektronix part number
103-0058-00
Checking Analog Bandwidth
22. Adapter SMA “T”, male to 2 SMA female Tektronix part number
015-1016-00
UsedtoTestDeltaTime
Measurement Accuracy
23. Adapter SMA female to BNC male Tektronix part number
015-0572-00
UsedtoTestDeltaTime
Measurement Accuracy
24. Adapter BNC male to female elbow Tektronix part number
103-0031-00
UsedtoTestDeltaTime
Measurement Accuracy
25. Adapter BNC female to clip lead Tektronix part number
013--0076--00
Used to Test Probe
Compensation Out put
26. Termination Short circuit, SMA connector Tektronix part number
015-1021-00
UsedtoTestDeltaTime
Measurement Accuracy
27. Attenuator, 2X Rati o: 2X; impedance 50 ; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0069-02
UsedtoTestDeltaTime
Measurement Accuracy
28. Mouse or keyboard Tektronix part numbers:
119-6298-xx (mouse)
119-6297-xx (keyboard)
Used to i nput test selections
1
Wavetek 9500/Option 100 and output head appropriate for the bandwidth of the oscilloscope under test (9520, 9530, 9550).
Warning: This generator can output dangerous voltages. Set the generator to Off or 0 volts before connecting, discon-
necting, or changing any test hookup during all procedures to follow. Also read the Warning statement on page 2-24.
2
For Del ta Time Measurement Accuracy, use a Wavetek 9500 or a pulse generator with a rise time as shown in Table 2-4
on page 2-44.
3
Warning: The P6243 and P6245 probes that may be used with this oscilloscope provide an extremely low loading
capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure
signals exceedi n g ±8 V, or errors in signal measurement will be observed. Above 40 V, dam age to the probe may result.
To make measurements beyond ±8 V, use either the P5050 probe (good to 500 V), or refer to the catalog for a recom-
mended probe.