Tektronix TDS5000B Series Sander User Manual


 
Performance Tests
2-60
TDS5000B Series Specifications and Performance Verification
H Adjust the output of the DC calibration generator until it precisely
overlaps the base (lower) level of the stored probe compensation
signal. (This value will be near zero volts.)
H Record the setting of the DC generator.
f. Press Close to remove the menus from the display. See Figure 2--27.
Figure 2- 27: Measurement of pr obe com pensator amplitude
g. Check against limits:
H Subtract the value just obtained (base level) from that obtained
previously (top level).
H CHECK that the difference obtained is within 990 mV to 1010 mV,
inclusive.
H Enter the voltage difference in test record.
3. Disconnect the hookup: Disconnect the cable from CH 1.
This completes the performance verification of the oscilloscope.