Tektronix TDS5000B Series Sander User Manual


 
TDS5000B Series Specifications and Performance Verification
1-1
Specifications
This chapter contains specifications for the TDS5000B Series Digital Phosphor
Oscilloscopes. All specifications are guaranteed unless labeled “typical.” Typical
specifications are provided for your convenience but are not guaranteed.
Specifications marked with the n symbol are verified in the Performance
Verification section.
The specifications in this section apply to all TDS5000B Series models unless
noted otherwise. To meet specifications, the following conditions must be met:
H The oscilloscope must have been calibrated in an ambient temperature
between 20 °C and 30 °C(68°F and 86 °F).
H The oscilloscope must be operating within the environmental limits listed in
Table 1-- 10 on page 1--22.
H The oscilloscope must be powered from a source that meets the specifica-
tions listed in Table 1--8 on page 1--21.
H The oscilloscope must have been operating continuously for at least 20
minutes within the specified operating temperature range.
H You must perform the Signal Path Compensation procedure after the
20-minute warm-up period and the ambient temperature must not change
more than 5 °C(9°F) without first repeating the procedure. See Run the
signal-path compensation routine on page 2--6 for instructions on how to
perform this procedure.
H You must perform the Signal Path Compensation procedure after the
20-minute warm-up period, and the ambient temperature must not change
more than 5 °C without first repeating the procedure. See Optimizing
Measurement Accuracy on page NO TAG for instructions to perform this
procedure.
Product and Feature Description
The TDS5000B Series Digital Phosphor Oscilloscope family consists of the
models shown in Table 1 --1.