Tektronix TDS5000B Series Sander User Manual


 
Performance Tests
TDS5000B Series Specifications and Performance Verification
2-61
Sine Wave Generator Leveling Procedure
Some procedures in this manual require a leveled sine wave generator to produce
the necessary test signals. If you do not have a leveled sine wave generator, use
one of the following procedures to level the output amplitude of your sine wave
generator.
If you have a power divider, use this setup to monitor the generator output
simultaneously with the level meter and the oscilloscope.
Equipment
required
Sine wave generator (Item 13)
Level met er and power sensor (Item 14)
Power di v ider (Item 15)
Two male N to female BNC adapters (Item 20)
One precision coaxi al cable (Item 4)
Prerequisites Read Prerequisit es on page 2--17 and foot note warnings on page 2--19.
1. Install the test hookup: Connect the equipment as shown in F igure 2-- 28.
TDS5000B Series oscilloscope
Sine wave
generator
Level meter
Power sensor
Power divider
Attenuators
(if necessary)
Input
Output
Figure 2- 28: Sine wave generator leveling equipment setup (Method A)
Method A