Tektronix TDS5000B Series Sander User Manual


 
Performance Tests
TDS5000B Series Specifications and Performance Verification
2-57
H CHECK that the readout CH 2 Pk-Pk is between 100 mV and
150 mV, inclusive.
H Enter the voltage in the test record.
4. Disconnect the hookup: Disconnect the cables from the channel inputs and
the rear panel outputs.
Equipment
required
One 10X oscilloscope probe, such as Tektronix P5050
Prerequisites See page 2--17. Also, the oscilloscope must have passed Check
Long-Term Sample Rate and Delay Time Accuracy on page 2--39.
1. Install the test hookup and preset the instrument controls:
a. Hook up test-signal: See Figure 2--24.
H Connect the probe to CH 1.
H Connect the probe leads to the PROBE COMP output.
TDS5000B Series oscilloscope
Passive probefrom PROBE COMP output
to CH1 input
Figure 2- 24: Initial test hookup
b. Initialize the oscilloscope: Press the DEFAULT SE TUP button.
c. Modify the initialized front-panel control settings:
H Set the Vertical SCALE to 200 mV.
H Set the Horizontal SCALE to 200 s.
H Press PUSH TO SET 50%.
H Use the Vertical POSITION knob to center the display on the
screen.
Check Probe
Compensation Output