Tektronix TDS5000B Series Sander User Manual


 
Performance Tests
TDS5000B Series Specifications and Performance Verification
2-51
stabilize the Delayed trigger. Click one of the S lope buttons to
switch between trigger slopes. See Figure 2--21 on page 2--50.
H Enter the pass/fail result for delayed trigger in the test record.
H Leave the Delayed trigger system triggered on the positive slope of
the waveform before continuing to the next step. Also, return to the
main trigger: select the A-->B Seq tab and click the A-->B Sequence
AOnlybutton.
H Select the AEventtab; then press Close.
3. Confirm the AUX Trigger input:
a. Display the test signal:
H Remove the 10X attenuator; then reconnect the 50 termination
and precision 50 cable to CH 1.
H Press the Trigger SOURCE button to select CH 1.
H Set the test signal amplitude for approximately 4 divisions on the
screen.
H Fine-adjust the generator output until the CH 1 Amplitude readout
indicates that the amplitude is 400 mV.
b. Check the AUX trigger source for stable triggering at 50 MHz: Do the
following in the order listed.
H Move the 50 termination and precision 50 cable from CH 1 to
AUX IN.
H Press the Trigger SOURCE button to set the source to EXT.
H Adjust the Trigger LEVEL knob in the positive direction.
H CHECK for a stable trigger. When a stable trigger is obtained, the
TRIG’D indicator is lighted.
NOTE. You will not see the signal on the display.
H Press the Trigger SLOPE button to toggle it to NE G.
H Adjust the Trigger LEVEL knob in the negative direction.
H CHECK for a stable trigger. When a stable trigger is obtained, the
TRIG’D indicator is lighted.
H Enter the pass/fail results in the test record.