Tektronix TDS5000B Series Sander User Manual


 
Brief Procedures
TDS5000B Series Specifications and Performance Verification
2-11
H Push the front-panel DELAY button several times to toggle delay off
and on and back off again. Verify that the display switches quickly
between two different points in time (the rising and falling edges of
this signal).
7. Remove the test hookup: Disconnect the probe from the channel input and
the probe compensation output.
Equipment
required
One 10X oscilloscope probe, such as Tektronix P5050
Prerequisites None
1. Initialize the oscilloscope: Push the front-panel DEFAULT SETUP button.
2. Hook up the signal source: Connect the probe to the probe compensation
output and to the C H 1 input as shown in Figure 2--5.
TDS5000B Series oscilloscope
Probe from PROBE
COMPENSATION
outputto CH 1input
Figure 2- 5: Setup for trigger test
3. Set up the oscilloscope: Push the front-panel AUTOSET button.
4. Verify that the main trigger system operates: Confirm that the following
statements are true.
H The trigger level readout for the A (main) trigger system changes with
the trigger-LEVEL knob.
H The trigger-LEVEL knob can trigger and untrigger the square-wave
signal as you rotate it. (Leave the signal untriggered).
H Pushing the front-panel trigger LEVEL knob sets the trigger level to the
50% amplitude point of the signal and triggers the signal that you just
left untriggered. (Leave the signal triggered.)
Verify t he A (Main) and B
(Delayed) Trigger Systems