Agilent Technologies E6385A Router User Manual


 
Performing Tests
Code Domain Analyzer Test
Chapter 4
O:\Manuals\E6385A_Cdma\Book\perf-tst.fm
72
Figure 4-9 Reading Individual Walsh Channel Measurements
When you have finished using the code domain analyzer screen, press the
Pause/Continue
key, and follow the directions on the screen. The Test Software will return to the Test
Selections Menu screen.
*
*
The Time and Phase measurements are only displayed if the
Measurement field is set to Phase or Timing. (See Figure 4-8).
To read the power Lvl, Time
error, or Phase error of an
individual Walsh Channel:
1) Set the Controls to Marker.
2) Set the Walsh Chan to the
channel to be tested.
3) Read the value(s).