Emerson 20/20 Air Compressor User Manual


 
14 – 6
Vission 20/20 • Operation and Service Manual •Vilter/Emerson • 35391SC 1.8.5153
Table 14-2. Slide Valve Troubleshooting Guide (2 of 2)
Problem Reason Solution
There is a rapid clicking noise when
the mot
or is operating. (Continued)
The position sensor’s EEPROM
memory has failed.
Replace the actuator.
There is a rapid clicking noise when
the motor is operating.
The photochopper is misaligned
with the slotted optocouplers.
Try to realign or replace the actuator.
The photochopper is positioned too
low on the motor shaft.
Adjust the photochopper so that
the fence extends further into the
optocoupler slots.
A motor bearing has failed. Replace the actuator.
The motor operates in one direction
only.
There is a loose connection in the
screw terminal blocks.
Tighten.
There is a loose or dirty connection
in the yellow Turck cable.
Clean and tighten.
The position sensor has failed. Replace the actuator.
There is a broken motor lead or
winding.
Replace the actuator.
The motor will not move in either
direction.
The thermal switch has tripped be-
cause the motor is overheated.
The motor will resume operation
when it cools. This could be caused
by a malfunctioning control panel.
Consult the factory.
Any of the reasons listed in “The mo-
tor operates in one direction only”.
See above.
The command shaft is jammed. Free the command shaft.
Broken gears in the gear-motor. Replace the actuator.
The motor runs intermittently,
several minutes on, several minutes
off.
Motor is overheating and the ther-
mal switch is tripping.
This could be caused by a malfunc-
tioning control panel. Consult the
factory.
The motor runs sporadically.
Bad thermal switch. Replace the actuator.
Any of the reasons listed in “The mo-
tor will not move in either direction”.
See above.
The motor runs but output shaft will
not turn.
Stripped gears inside the gear rotor,
or the armature has come unpressed
from the armature shaft.
Replace the actuator.
Section 14 Slide Calibration