Emerson HASAxE-IM-HS Welding Consumables User Manual


 
Instruction Manual
HASAxE-IM-HS
04/2006
X-STREAM
Emerson Process Management GmbH & Co. OHG S-9
Safety Instructions
General Operating Instructions
DANGER TO LIFE ! EXPLOSION HAZARD !
Verify all gas lines are connected as described within this manual and tight!
Improper gas connections may cause explosion, serious injury or death!
Exhaust may contain hydrocarbons and other toxic gases, e.g. carbon
monoxide. Carbon monoxide is toxic!
Indoor installation area has to be clean, free from moisture, excessive vibration and frost-
protected.
Take care to meet the permissible ambient temperatures as given in the technical data section!
Instruments must not be exposed to direct sunlight nor sources of heat. Do not cover venting
openings and take care to mount the instrument in a distance to walls not affecting venting.
Do not interchange gas inlet and outlet! All gases must be conditioned before supplying! When
supplying corrosive gases ensure that gas path components are not affected!
Max. permissible gas pressure: 1,500 hPa (7.5 psig), except with paramagnetic Oxygen sensor
(atmospheric pressure;
page 3-17)!
Exhaust lines must be installed in a descending way, need to be pressureless, frost-protected
and in compliance with applicable legislative requirements!
When it is necessary to open gas paths seal the analyzer‘s gas fittings by using PVC caps to
avoid pollution of the internal gas path by moisture, dust, etc.
To stay in compliance with regulations regarding electromagnetic compatibility it is recommended
to use only shielded cables, as optionally available from Emerson Process Management or
equivalent. Customer has to take care that the shield is connected in proper way ( section
4-5, page 4-29). Shield and signal connector enclosure need to be conductively connected,
submin-d plugs and sockets must be screwed to the analyzer.
Using external submin-d-to-terminal adaptor elements (option) affects electromagnetic
compatibility. In this case the customer has to take measures to stay in compliance and has to
declare conformity, when required by legislation (e.g. European EMC Directive).