6-2 Checking the Instrument's Setup
Next check the display settings for the mea-
surement screen, temperature & pressure units
and menu access using the DOWN and ENTER
keys to access the related menu pages.
If system is setup accordingly access level 2
code must be entered to access these menus.
Factory default setting is "no code required".
If any of the settings do not meet your require-
ments just enter the related menu and change
the parameter.
For easier handling it is recommended to not
change the menu access codes settings during
the first instrument setup check.
If system is setup accordingly access level 2
code must be entered to access these menus.
Factory default setting is "no code required".
Dual channel instrument:
Optional gas component selection menu -
Select the component to be setup
After the display settings are checked press
the LEFT key to return to the SETUP menu to
enter the CALIBRATION setup menu for
specifying the calibration gas concentrations
and more.
Calibration..
Calibration gases..
Within this screen enter the values for the listed
gases: Take the correct values from your gas
supplier's certificate. Proper values only assure
faultless measuring results.
For multichannel instruments enter the values
for each channel separately.
Calibration gases
ZeroGas: 0.0 ppm
SpanGas: 500.0 ppm
Press the LEFT key to return to the DISPLAY
SETUP screen.
6-2-2 Display Information Setup
6-2-3 Calibration Data Setup