X-STREAM
2-2
Instruction Manual
HASAxE-IM-HS
05/2006
Emerson Process Management GmbH & Co. OHG
2-1 Common Technical Data
2-1 Common Technical Data
Site of installation
Humidity (non condensing) < 90 % r. h. at 68 F (+20 °C)
< 70 % r. h. at 104 F (+40 °C)
Pollution degree 2
Installation category I I
Altitude 0 to 6560 ft (2000 m) above sea level
Sourrounding atmosphere Analyzers must not be operated in hazar-
dous or flammable atmoshere without
additional safety measures.
Analyzers must not be operated in
corrosive atmoshere.
Power supply
Rated input voltage 100 - 240 V
50/60 Hz, wide range input
Power supply voltage fluctuations are
not to exceed +/- 10 % of the nominal
supply voltage!
Input voltage range 85 - 264 V , 47 - 63 Hz
Rated input current
standard 0,75 - 0,35 A max.
with thermostated physics 2 - 1 A max.
Compliances
Electrical safety CAN / USA CSA-C/US, based on
CAN/CSA-C22.2 No. 61010-1-04 /
UL 61010-1, 2nd Edition
Europe CE, based on EN 61010-1
Electromagnetic compatibility
Europe CE, based on EN 61326
Australia C-Tick
other NAMUR