Emerson Process Management GmbH & Co. OHG7-14
Instruction Manual
HASAxE-IM-HS
05/2006
7-3-3 Advanced Calibration
To start a ZERO calibration for ALL channels
select the second line.
Starting from the measurement screen press
the DOWN key to open the MAIN MENU and
enter the CONTROL - ADV.CALIBRATION
menu.
The analyzer immediately begins zero
calibration(s). Watch the screen for status
information (explained by means of exemplary
pictures on the next page).
7-3-3-2 Zero All Calibration
Before selecting any further line
make sure the required calibra-
tion gas is applied!
Supply all calibration gases with
the same flow as the sample gas
(recommeded approx. 1 l/min),
pressureless and utilizing the
right gas fitting (see sect. 3-4).
Make sure the calibration purge
time is set to a value ensuring the
measuring cell is filled pro-perly
with the related calibration gas
after the valve has opened!
Ensure the warm-up time after
switching on has elapsed!
Warm-up time is 15 to 50 minutes
depending on installed measu-
ring system and configuration!
Cancel calibration!
ZeroAll!
SpanAll!
ZeroSpanAll!
Control..
Adv.Calibration..
Note!
Single channel analyzers show the same
menu, with the restriction, that the term "ALL"
relates to the single channel only!