7-5 Cleaning the Instrument's Outside
7-5 Cleaning the Instrument's Outside
Do NOT drench the cloth, just
moisten it to prevent liquid ente-
ring the housing!
Clean the analyzer housing outside with the
moistened cloth.
If required dry the housing after cleaning.
Use a liquid general purpose detergent and a
lint-free cloth for cleaning the analyzer´s out-
side.
Procedure
Disconnect instrument from mains!
Before opening gas paths they
must be purged with ambient
air or neutral gas (N
2
) to avoid
hazards caused by toxic, flam-
mable, explosive or harmful to
health sample gas compo-
nents!
If opening the gas paths is
required, seal the open
analyzer´s gas fittings utilizing
PVC caps to avoid pollution of
inner gas path.
Moisten the lint-free cloth with a mixture of
3 parts of water and 1 part of the general
purpose detergent.