9 Modbus
9-3 Modbus - List of Parameters and Registers
This list is ordered alphabetically by the parameter name („Mnemonic“) column.
Any "#" is a wildcard, to be replaced by "1" or "2" for the desired measuring channel. In this case
the upper register line is assigned to channel 1, the lower line to channel 2.
Attribute
1
Daniel /
Enron
Modicon Type
Range / Enum values / Coefficients
Modbus-Register
Mnemonic
Description
3061..3065 3061..3065
SerialNumberDB
serial number
2 characters per register:
1
st
character at HIGH byte,
2
nd
character at LOW b
y
te.
Cfg, Fac
String-10
1020 1020
SimulationDB
any kind of value simulation is running
(ActivateSimulValue#DB,
ActivSimulMuxValuesDB)
Dyn
Boolean
3057 3057
SIntInstalledDB
Defines whether serial interface hardware is
installed
0 = not installed
1 = installed
Cfg, Fac
Boolean
3058 3058
SIntModbusFt32DB
In Modbus-Protocol:
transmission format of 32-bit registers
0 = 32bit (Daniel / Enron)
1 = 16bit low word first (default
Modicon)
2 = 16bit high word first (swapped
Modicon
)
Cfg
Enum
9029
9030
6057..6058
6059..6060
SourceCurrent#DB
calculated current of light source in mA
Dyn
Float
9011
9012
6021..6022
6023..6024
SpanFactor#DB
span correction factor (calculated by span
calibration)
Cfg, Self
Float
9037
9038
6073..6074
6075..6076
SpanGasValue#DB
value which a span calibration adjusts to
unit is ppm;
min/max depend on EndOfRange#DB,
MaxConcePercent#DB,
DifferentialMeasurementDB
Cfg
Float
1011
1012
1011
1012
SpanGasValueIncorrect#DB
spangas = zerogas value not allowed
Dyn
Boolean
1007
1008
1007
1008
SpanTolerance#DB
tolerance check did not allow span calibration
Dyn
Boolean
3121..3124 3121..3124
SpecialAccessCodeDB
user code for getting access to specialist areas
2 characters per register:
1
st
character at HIGH byte,
2
nd
character at LOW byte.
Cfg
String-8
3125 3125
SpecialAccModeDB
mode for getting access to specialist areas
0 = access is allowed
1 = access requires user code
2 = access is prohibited
Cfg
Enum
9031
9032
6061..6062
6063..6064
StartOfRange#DB
beginning of measurement range in ppm
0
Rd
Float
9085
9086
6169..6170
6171..6172
TcntCorrCoeff#DB
gradient m of equation 'RawValueTfact#DB =
m*Temperature + 1'; see RawValueTfact#DB
Cfg, Fac
Float