Emerson HASAxE-IM-HS Welding Consumables User Manual


 
X-STREAM
Emerson Process Management GmbH & Co. OHG7-46
Instruction Manual
HASAxE-IM-HS
05/2006
If system is setup accordingly access level 3
code must be entered to access this menu.
7-6 Save / Restore Configuration Data Sets
Starting from the measurement screen press
the DOWN key to open the MAIN MENU, enter
the SETUP and next the SAVE-LOAD.. menu.
7-6-1 Save CfgData to UserData
2
nd
menu page
FactData>CfgData..
CfgData>UserData..
UserData>CfgData..
Now select the "CfgData>UserData.." line and
press the ENTER key.
Save-Load
CfgData>COMport!
COMport>CfgData..
Verify!
1
st
menu page
Press the DOWN key to open the second menu
page.
CfgData>UserData
Are you sure?
No!
Yes!
A screen appears to confirm the operation:
Select Yes! and after pressing the ENTER key
a new screen shows up showing the current
status.
Copying data
- PLEASE WAIT -
Procedure X:078000
The instrument now stores the currently used
(and changed by operator) analyzer setup into
a special memory area. This data is then called
UserData and used for backup only, while the
data used for operation is called CfgData.
Any further changes affecting the instrument's
setup update the CfgData only, as long as not
again stored into the UserData set.
When the procedure has finished, the
COMMAND EXECUTED screen shows up.
(i)
-COMMAND EXECUTED-
Setup..
Save-Load..