X-STREAM
5-31
Instruction Manual
HASAxE-IM-HS
05/2006
Emerson Process Management GmbH & Co. OHG
5 User Interface
5-4 Menu System - Measurement Setup Menu
5-4-3-3-2 Cross Interference Setup Menu
Toggle the status for cross interference
compensation:
Available options:
Off: Cross compensation inactive
On: Two-way cross compensation is calcu-
lated for both channels.
If system is setup accordingly access level 3
code must be entered to access this menu.
CrossInterference
Status: Off
@ Span Cal: No
Toggle the status for cross interference
compensation during span calibration:
Available options:
No: disabled
Yes: Enabled - Cross compensation factors
are calculated during span calibration. Requires
using pure gas for span calibration!
Note!
Setting this parameter to Yes is valid for the
next span calibration only and will be reset to
No automatically after calibration has
finished.
Note!
This menu is available only for dual channel
instruments.
Setup..
Measurement..
CrossInterference..