Emerson Process Management GmbH & Co. OHG6-20
Instruction Manual
HASAxE-IM-HS
05/2006
Press the HOME key to return to the
MEASURING SCREEN, because the setup
checking procedure is finished.
6-2 Checking the Instrument's Setup
CfgData>UserData
Are you sure?
No!
Yes!
A screen appears to confirm the operation:
Select Yes! and after pressing the ENTER key
a new screen shows up showing the current
status.
Copying data
- PLEASE WAIT -
Procedure X:078000
The instrument now stores the currently used
(and changed by operator) analyzer setup into
a special memory area. This data is then called
UserData and used for backup only, while the
data used for operation is called CfgData.
Any further changes affecting the instrument's
setup update the CfgData only, as long as not
again stored into the UserData set.
Other options in the SAVE-LOAD-UPDATE
menu allow to restore the UserData into the
CfgData, in case the analyzer setup has been
changed into a undesired status.
When the procedure has finished, the
COMMAND EXECUTED screen shows up.
Note!
For a detailed description of all SAVE-LOAD-
UPDATE options
7-6, page 7-34.
(i)
-COMMAND EXECUTED-