POWER SUPPLY
Verify the mains voltage at site of installation corresponds to the analyzer´s
rated voltage as given on the nameplate label!
Verify the safety instruction given by power supply unit manufacturer !
Installation and connecting mains and signal cables are subject to qualified
personnel only taking into account all applicable standards and legislative
requirements!
Failure to follow may cause warranty invalidation, property damage and/or
personal injury or death! Connecting mains and signal cables to internal
srew terminals requires working at open housing near life parts!
Installation of this instrument is subject to qualified personnel only, familiar
with the resulting potential risks!
The gas analyzers do not provide a mains power switch and are operable
when connected to power.
The gas analyzers do not provide a mains switch! A mains switch or circuit
breaker (to comply with IEC 60947-1 /-3) has to be provided in the building
installation. This switch has to be installed near by analyzer, must be easily
operator accessible and has to be assigned as disconnector for the analyzer.
Cables for external data processing must be double insulated for mains
voltage when used inside the instrument!
If double insulation is not available signal cables inside the analyzer must
be installed in a way that a distance of at least 5 mm is ensured permanently
(e.g. by utilizing cable ties).